341
Micro-cell Electron Capture Detector (
µ
-ECD)
Replacement procedures
14 of 26
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 341-10
Bracket mounted onto the manifold
3.
Peel the blank label from its backing and paste it on the mounting bracket
over the screw heads.
4.
If the tubing from the gang fitting bends to the left, reshape it so that it
bends up and back from the block as shown in Figure 341-12.
Finger-tighten nuts.
Do not use a wrench.
(FPD shown above)
7/16-inch hex nut
Mounting bracket
ID tag through
Ribbon cable
Gang fitting installs here
Cover screw
slot in bracket
blank label
heads with