220
Purged/Packed Inlet
Replacement procedures
6 of 20
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
8.
Lift the inlet out of the instrument.
Figure 220-6
Removing the top inlet plate
9.
If you need to remove the thermal block, slide off the insulation sleeve
and use a wrench to remove the bottom nut. Slide the thermal block off
the inlet weldment.