Agilent 6890 Gas Chromatograph Service Manual
Flame Photometric Detector (FPD)
350
Replacement procedures
Figure 350-14
Removing an FPD flow manifold
Caution
Always hold the Type 2 manifold by its support bracket to avoid damaging
board components.
7.
Remove the manifold from its slot.
Installing a Type 2 manifold
WARNING
Hydrogen gas is flammable and potentially explosive. Before replacing the
manifold, turn off the hydrogen gas at the source.
Remove the screw
Type 1 manifold
Type 2 manifold