360
Auxiliary EPC
Replacement procedures
12 of 16
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
15. Attach the Aux EPC label to the top of the pneumatics chassis as shown
in Figure 360-9 below.
Figure 360-9
Attaching the Aux EPC label
16. Install the RFI shield, the pneumatics cover, and the detector top cover.
17. If using hydrogen or other flammable gases, install an external flow
restrictor, part no. G1530-80380, on each fitting that will be used for a
flammable gas.
18. Check for leaks.
Restore the GC to operating condition
1.
Plug in the GC and turn it on.
Gas outlet block
Attach label here
External flow restrictor
installed and ready for
connection to gas line.