Programmed Temperature Vaporization Inlet (PTV)
Solvent vent mode
12 of 46
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Stage 3: Purge and cleanup
The solenoid valve opens again and the system returns to the Stage 1
configuration but with different setpoints. The PTV inlet is flushed through
the split vent.
Figure 240-6
Solvent vent mode flow control diagram: purge and cleanup
Split
vent
SPR
vent
Septum
holder
To detector
Flow
limiting
frit
Total flow
control loop
PS
SPR
PV1
PV2
FS
Cryo
valve
PTV thermocouple PCB
Trap
Valve
open