Flame Photometric Detector (FPD)
Replacement procedures
26 of 36
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off all detector gases at their supply, then turn off the main power
switch and unplug the power cord.
Caution
Always hold the manifold as shown below to avoid damaging board
components.
1.
While holding the manifold by the black plastic frame as shown below,
slip the label through the slot in the mounting bracket, and align the
bracket holes over the gas fittings.
2.
Finger-tighten three 7/16-inch hex nuts over the fittings to hold the
bracket in place. It is
very important
that you do
not
tighten the nuts yet.
Hold manifold by
the black plastic frame.
Slip label tag through
the slot in the bracket.