420
Oven and Temperature Control
Replacing the oven flap motor
18 of 20
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 420-10
Removing the flapper assembly
4.
Remove the two Torx T-10 screws securing the motor and flapper
assembly to the bracket and remove the motor/flappers from the bracket.
5.
Remove the Torx T-20 screw at the bottom of the flapper shaft near the
flap motor and disconnect the motor from the flappers.
6.
Reassembly is the reverse of removal.
Shroud
Motor
Flappers