Agilent 6890 Gas Chromatograph Service Manual
Covers and Fans
410
Removing the ALS tray bracket
Removing the ALS tray bracket
When a 7673 Automatic Liquid Sampler is installed on a 6890 GC (serial
number < 20,000), a large plastic tray bracket replaces the blue inlet carrier
cover. Follow the procedure below to remove this bracket.
1.
Remove the two screws located on the under side of the rounded end of
the tray bracket assembly using a T-20 Torx driver.
2.
Remove the two screws at either end of the tray mounting strap using a
T-20 Torx driver.
3.
Remove six screws attaching the bracket assembly to the inlet chassis
using a T-20 Torx driver. Lift the assembly straight up.
Figure 410-11
Removing the ALS tray bracket
Tray bracket assembly
Screw (2)
M4 ×10 PH Torx
GC mainframe
Left side cover
Tray mounting strap
Screw (2), M4 ×18 FH Torx
Screws (6)