Programmed Temperature Vaporization Inlet (PTV)
Replacement procedures
32 of 46
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
9.
Check the entire system again for leaks; if necessary, carefully retighten
the guide cap with the syringe needle inserted slightly more and/or
replace the Kalrez seal.
If the head leaks when a syringe is inserted, the Teflon ferrule is the
problem. If the head leaks without a syringe inserted, the seals may need
to be replaced.
Replacing the Teflon ferrule
1.
Unscrew the guide cap from the septumless head and remove the Teflon
ferrule.
Figure 240-21
Installing a Teflon ferrule