Checkout Chromatograms
FPD checkout conditions and chromatogram
22 of 26
Jun 2001
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
FPD checkout conditions and chromatogram
Column and sample
Type
HP-5 30m × 0.32 mm × 0.25 µm PN 19091J-413
Sample
FPD Checkout PN 8500-3697
Injection volume
1 µL
Inlet
Temperature
250°C Purged/Packed
250°C
Split/Splitless
Oven TrackCool On-Column
80°C
PTV (see below)
Inlet pressure
25 psi (Constant pressure for EPC inlets,
helium)
Split/Splitless Only
Mode
Splitless
Purge flow
60 mL/min
Purge time
0.75 min
PTV—Splitless Mode
Inlet temperature
80°C
Initial time
0.1 min
Rate 1
720°C/min
Final temp 1
350°C
Final time 1
2 min
Rate 2
100°C/min
Final temp 2
250°C
Final time 2
0 min
Inlet pressure
25 psi (const. press. for EPC inlets)
Purge time
0.75 min
Purge flow
60 mL/min