Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
310
Replacement procedures
8.
Loosen the four captive Torx T-20 screws from the top of the detector
pallet and remove the detector from the instrument.
Figure 310-6
Removing the detector
Four captive screws
Ignitor cable
Electrometer ribbon
cable
Detector pallet