Agilent 6890 Gas Chromatograph Service Manual
Thermal Conductivity Detector (TCD)
330
Theory of operation
EPC detector
The EPC TCD has one flow line for the makeup/reference gas. After an initial
filter frit, it splits into two paths. Each line has a proportional valve controlled
by a pressure sensor, and a non-adjustable restrictor frit. In addition, the
switching gas line has a switching valve that switches paths through the TCD
cell several times a second. The switching line oscillates between two
flowpaths, one for each channel of the TCD cell. The makeup gas line sweeps
past the end of the column and carries the column effluent into the TCD cell.
Figure 330-4
EPC TCD flow diagram
Manually controlled detector
The manually controlled TCD has a pneumatic supply line for the switching
gas (also called modulating or reference gas) flow and one for the makeup
gas. Both of the lines have pressure regulators, an on/off solenoid valve and
a non-adjustable restrictor frit. In addition, the switching gas line has a
Filter
frit
Makeup and
reference gas in
Proportional
valves
Pressure
sensors
Restrictors
Reference
switching
valve
Vent
Makeup flow
PS
PS
Pressure
control loops
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......