Agilent 6890 Gas Chromatograph Service Manual
Volatiles Interface
250
Replacement procedures
WARNING
Turn off the oven and the interface and allow them to cool. Turn off all flows
at the initial gas supply. Then turn off the main power switch and unplug the
power cord.
1.
Disconnect the transfer line. Loosen the nut with a 1/4-inch wrench and
remove the line. Remove the clamping plate from the interface by
loosening the captive screw with a T-20 Torx screwdriver. Put the plate
in a safe place.
Figure 250-6
Removing the transfer line
2.
Lift the interface out of the heater block.
Transfer line
Clamping plate