Gas requirements
2 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Gases for capillary columns
When used with capillary columns, GC detectors require a separate makeup
gas for optimum sensitivity. For each detector and carrier gas, there is a
preferred choice for makeup gas. Table 520-2 lists gas recommendations for
capillary columns.
Table 520-2
Gas Recommendations for Capillary Columns
Detector
Carrier
gas
Preferred
makeup gas
Second choice
Detector anode
purge or reference
gas
Electron Capture
Hydrogen
Argon/Methane
Nitrogen
Anode purge must
be same as makeup
Helium
Argon/Methane
Nitrogen
Nitrogen
Nitrogen
Argon/Methane
Argon/
Methane
Argon/Methane
Nitrogen
Flame Ionization
Hydrogen
Nitrogen
Helium
Hydrogen and
air for detector
Helium
Nitrogen
Helium
Nitrogen
Nitrogen
Helium
Flame
Photometric
Hydrogen
Nitrogen
Hydrogen and air for
detector
Helium
Nitrogen
Nitrogen
Nitrogen
Argon
Nitrogen
Nitrogen-
Phosphorus
Helium
Nitrogen
Helium**
Hydrogen and
air for detector
Nitrogen
Nitrogen
Helium**
Thermal
Conductivity
Hydrogen*
Must be same
as carrier and
reference gas
Must be same
as carrier and
reference gas
Reference must be
same as carrier and
makeup
Helium
Nitrogen
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......