1015
Cabling Diagrams, 6890N
6890N GC GC Automatic Liquid Sampler Non-Agilent Data System
4 of 14
Jun 2001
Cabling Configuration and Electronics
Agilent 6890 Gas Chromatograph Service Manual
6890N GC
GC Automatic Liquid Sampler
Non-Agilent Data System
Number
Part no. and description
1
35900-60670, General use APG remote cable, 9-pin male/spade lug (2m)
35900-60920, General use APG remote cable, 9-pin male/spade lug (5m)
35900-60930, General use APG remote cable, 9-pin male/spade lug (0.5m)
2
G1530-60590, External event cable, 8-pin/spade lugs
35900-60670 APG remote
cable spade lug identification
G1530-60590 External event
cable spade lug identification
Connect 1
9 pin
(male)
Signal name
Connector 2
spade lugs
Pin
Color
Signal
1
GND
Black
1
Yellow
24 V Out 1
2
Prepare
White
2
Black
24 V Out 2
3
Start
Red
3
Red
Ground
4
Shut down
Green
4
White
Ground
5
Reserved
Brown
5
Orange
Contact 1
6
Power on
Blue
6
Green
Contact 1
7
Ready
Orange
7
Brown
Contact 2
8
Stop
Yellow
8
Blue
Contact 2
9
Start Request
Violet
The 7683 ALS
controller is internal to
the 6890N GC.
The G2613A injector
and G2614A tray
plug into the GC.
6890N GC
1 AGP remote
2 External event
Spade lug termination
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......