Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
860
Bad Mainboard and Fatal Errors
These messages almost always indicate that the mainboard is malfunctioning
and must be replaced. These messages are not numbered and usually appear
when the instrument is first turned on. Refer to the following table for a list
of messages.
Table 860-1
Bad Mainboard and Fatal Error Messages
Pop-up message
Comments
BAD MAINBOARD
Main FPGA Failure
Replace mainboard
Static RAM Failure
Replace mainboard
Boot ROM Checksum
Replace mainboard
ROM #2 Checksum
Replace EEPROM 2.
ROM #3 Checksum
Replace EEPROM 3.
Incorrect ROM #2
EEPROM 2 or 3 is installed in the wrong position. Change the
position the EEPROM is installed in.
Incorrect ROM #3
ROM #2 wrong version
The version of either EEPROM 2 or 3 is different than that of
the other EEPROMs.
ROM #3 wrong version
DMA FPGA Failure
Replace mainboard
DRAM Failure
Replace mainboard
FATAL ERROR
Replace mainboard for all of the following errors.
Exception Vector
Bus Error
Address Error
Illegal Instruction
Divide by Zero
No 512Hz Interrupt
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......