1 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
820
Not Ready Messages
This section lists Not Ready messages according to the type of condition the
GC has detected. The GC does not provide a popup message when a component
is not ready; instead, the not ready LED blinks and the run simply does not
begin. The messages discussed in this section correspond to those you see
when you press the
[
Status]
key. Not Ready messages fall into these
categories:
•
Oven temperature not ready
•
Temperature zones not ready
•
Pressure or flow not ready
•
Detector not ready
•
GC in diagnostic mode
•
Inlet purging
•
Valves not ready
•
External devices not ready
•
Power failure
Oven temperature not ready
Status message: Oven temp
The GC will becomes ready to begin a run when the oven temperature remains
within 1°C of the setpoint for the equilibration time. The oven must be turned
on.
❐
The instrument is not ready until the oven is turned on and reaches its
setpoint. If the oven is unable to reach the setpoint, the GC will remain
not ready indefinitely unless the oven temperature is out of the oven
range, which will cause a shutdown.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......