210
Split/Splitless Inlet
EPC inlet
4 of 30
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 210-2
EPC control splitless mode flow diagram (pre-run to purge time).
Pulsed splitless mode
For EPC inlets set to pulsed splitless mode, the column head pressure is
increased to some selected value and selected time period prior to sample
introduction. Like the split mode, the increased flow through the column
during the injection period minimizes sample losses in glass liners.
Gas saver mode
For EPC inlets, Gas Saver mode may be used in splitless or split mode of
operation for conservation of split and inlet purge flows when not in use or
in between runs. A gas saver flow rate and start time can be set independently
of normal flow rates being used.
Flow
limiting
frit
PV1
Proportional
valve 1
Flow
sensor
Septum holder
Pressure
sensor
Septum purge
regulator (not
adjustable)
Septum purge
3 mL/min
Inlet vent (split)
PV2
Proportional
valve 2
Valve
closed
1 m
L/
m
in
1m
L/
m
in
PS
SPR
FS
Safety shutdown mode:
Proportional valve 1 closed
Proportional valve 2 open
On/off valve open
V
4 mL/min
To Detector
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......