7 of 10
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
External Connectors, 6890N
1215
Remote start/stop
Remote start/stop
Figure 1215-6
Remote start/stop connection
Signal descriptions
Prepare (low)
—
Request to prepare for analysis. Receiver is any module
performing pre-analysis activities.
Start (low)
—
Request to start run/timetable. Receiver is any module
performing runtime-controlled activities.
Ready (high)
—
System is ready for next analysis. Receiver is any sequence
controller.
Stop (low)
—
Request to reach system ready state as soon as possible (for
example, stop run, abort or finish, and stop injection). Receiver is any module
performing runtime-controlled activities.
S
A
M
P
L
E
R
2
T
R
A
Y
S
I
G
1
S
I
G
2
E
V
E
N
T
B
C
D
R
S
/
2
3
2
R
E
M
O
T
E
LAN
S
A
M
P
L
E
R
1
S
A
M
P
L
E
R
1
S
I
G
1
R
E
M
O
T
E
1
6
5
9
JP1, REMOTE
Pin
Function
Logic
1
Digital ground
2
Prepare
LOW true
3
Start
LOW true (input)
4
Start relay
5
Start relay
6
No connection
7
Ready
HIGH true (output)
8
Stop
LOW true
9
No connection
REMOTE
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......