CI Troubleshooting
3
75
No peaks
When troubleshooting no peaks, it is important to specify what mode of
operation is being used, and what expected peaks are not being seen. Always
start with methane PCI and verify presence of reagent ions.
No reagent gas peaks in PCI
If the MSD has been working well and nothing seems to have been changed
•
Wrong tune file loaded, or tune file corrupted
•
Wrong ion polarity (there are no reagent ions visible in NCI)
•
No reagent gas flow; look for background ions, and check pressure
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
•
Large air leak
•
Dirty ion source
•
Poor vacuum (pump problem), see
.
If the MSD was recently switched from EI to CI
•
Isolation tip not installed
•
No reagent gas flow
•
Analyzer not sealed (big air leak)
•
Wrong tune file loaded or tune file corrupted
•
Ion source not assembled or connected correctly
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
No PFDTD peaks in PCI
•
Incorrect reagent gas. There are
no
PCI PFDTD peaks created with
isobutane or ammonia. Switch to methane.
•
Analyzer not sealed (big air leak)
•
No calibrant in vial
•
Defective calibration valve(s)
•
Air leak in carrier or reagent gas path
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......