122
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
8
Pour new DP fluid into the DP until the fluid level is within the FULL
COLD range.
The recommended charge for this pump is 30 mL. It will require
approximately 1.5 of the bottles (18.5 mL each) of DP fluid. Pour the fluid
between the center stack and the side wall. Watch the sight glass while
pouring. Do not overfill.
9
Reinstall the DP. (See