208
5977B Series MSD Troubleshooting and Maintenance Manual
7
Analyzer
Analyzer Heaters and Radiators
The ion source and mass filter are housed in cylindrical aluminum tubes
called radiators (See
The radiators control the distribution of heat in the analyzer. They also
provide electrical shielding for analyzer components. The source heater and
temperature sensor are mounted in the source heater block. The mass filter
(quad) heater and temperature sensor are mounted on the mass filter
radiator. Analyzer temperatures can be set and monitored from the MSD Data
Acquisition software.
In selecting the temperatures to use, consider the following:
•
Higher temperatures help keep the analyzer clean longer.
•
Higher ion source temperatures result in more fragmentation and,
therefore, lower high-mass sensitivity.
After pumpdown, it takes at least 2 hours for the analyzer to reach thermal
equilibrium. Data acquired sooner may not be reproducible.
Recommended settings (for EI operation):
•
Ion source 230
C
•
Quadrupole 150
C
The GC/MSD interface, ion source, and mass filter (quad) heated zones
interact. The analyzer heaters may not be able to accurately control
temperatures if the setpoint for one zone is much lower than that of an
adjacent zone.
C A U T I O N
Do not exceed 200 °C on the quadrupole, or 350 °C on the ion source.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......