General Troubleshooting
2
5977B Series MSD Troubleshooting and Maintenance Manual
41
•
Very dirty ion source
•
Calibration valve is not working correctly
•
Bad signal cable connection
•
Filament has failed, or is not connected correctly
•
Bad ion source wiring connection
•
Bad detector wiring connection
•
Failed electron multiplier horn
Peaks are tailing
•
Active sites in the sample path
•
Injection is too large
•
Incorrect GC inlet temperature
•
Insufficient column flow
•
GC/MSD interface temperature is too low
•
Ion source temperature is too low
Peaks are fronting
•
Column film thickness mismatched with analyte concentration
(column overload)
•
Initial oven temperature is too low
•
Active sites in the sample path
•
Injection is too large
•
GC inlet pressure too high
•
Insufficient column flow
Peaks have flat tops
•
Insufficient solvent delay
•
Incorrect scale on the display
•
Injection is too large
•
Electron multiplier voltage is too high
•
Gain is too high
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......