General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
129
9
Withdraw the vial 1 mm. This prevents damage when you tighten the
collar.
10
Turn the collar clockwise to tighten it.
The collar should be snug but not overly tight. Do
not
use a tool to tighten
the collar. It does not require that much force.
11
Reinstall the analyzer window cover.
12
Purge the EI calibration valve. (See
“To Purge the Calibration Valves”
C A U T I O N
Failure to purge the calibration valve will result in damage to the filaments and
detector.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......