General Troubleshooting
Poor sensitivity
•
Incorrect tuning
•
Tune file that does not match the type of analysis
•
Repeller voltage is too low
•
Incorrect temperatures (oven, GC/MSD interface, ion source, or mass filter)
•
Incorrect sample concentration
•
Leaking GC inlet
†
•
Dirty GC inlet
•
Incorrect split ratio
•
Purge off time in splitless mode is too short
•
Excessive pressure in the MSD
•
Dirty ion source
•
Air leak
•
Poor filament operation
•
Detector (HED electron multiplier) is not working correctly
•
Incorrect mass filter polarity
† This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
•
Dirty syringe needle
•
Dirty GC inlet
•
Leaking GC inlet
†
•
Injection is too large
•
Loose column connections
•
Variations in pressure, column flow, and temperature
•
Dirty ion source
•
Loose connections in the analyzer
•
Ground loops
† This could cause a fault condition in the GC that would prevent the GC
from operating.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......