70
3
CI Troubleshooting
Pressure-Related Symptoms
The following symptoms are all related to high vacuum pressure. Each
symptom is discussed in more detail in the following pages.
The mass flow controller is calibrated for methane, and the high vacuum gauge
controller is calibrated for nitrogen, so these measurements are not accurate
in absolute terms (
). They are intended as a guide to typical observed
readings. They were taken with the following set of conditions:
Source temperature
250 °C
Quad temperature
150 °C
Interface temperature
280 °C
Helium carrier gas flow
1 mL/min
Table 8
Pressure measurements
Pressure (Torr)
MFC (%)
Methane
Ammonia
10
5.5
×
10
–5
5.0
×
10
–5
15
8.0
×
10
–5
7.0
×
10
–5
20
1.0
×
10
–4
8.5
×
10
–5
25
1.2
×
10
–4
1.0
×
10
–4
30
1.5
×
10
–4
1.2
×
10
–4
35
2.0
×
10
–4
1.5
×
10
–4
40
2.5
×
10
–4
2.0
×
10
–4
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......