5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Install the EI Calibration and Vent Valve Assembly
Materials needed
•
Calibration valve
•
Diffusion (G7077-60211)
•
Turbo (G7077-60204)
•
O-ring for calibration valve (0905-1217)
•
PFTBA (05971-60571) or other tuning compound
•
Screwdriver, Torx T-20 (8710-1615)
Procedure
1
Remove the old valve assembly. (See
“To Remove the EI Calibration and
2
Ensure the valve O-ring is in place. If it is worn or damaged, replace it.
3
Install the calibration and vent valve assembly, and tighten the screws that
hold it in place.
4
Reconnect the calibration valve cable to the connector next to the fan.
5
Remove the vial from the new calibration valve. (See
on page 128.) The valve is supplied with a vial already
installed.
6
Fill and reinstall the calibration vial.
7
Pump down the MSD.
C A U T I O N
Failure to purge the calibration valve will damage the filaments and detector.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......