![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 64](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608064.webp)
64
3
CI Troubleshooting
Common CI-Specific Problems
Because of the added complexity of the parts required for CI, there are many
potential problems added. The greatest number of, and most serious problems
with CI are associated with leaks or contamination in the reagent gas
introduction system. NCI is especially sensitive to the presence of air; small
leaks that cause no problems in PCI can destroy NCI sensitivity.
As with EI, if the MSD tunes well and no air leak is present, sample sensitivity
problems should be addressed by GC inlet maintenance first.
•
Wrong reagent gas
•
Reagent gas not hooked up or hooked up to wrong reagent gas inlet port
•
Wrong ions entered in tune file
•
Wrong tune file selected
•
Not enough bakeout time has elapsed since vent (background is too high)
•
Wrong column positioning (extending > 2 mm past tip of interface)
•
Interface tip seal not installed
•
EI ion source installed in CI mode
•
EI filament or other EI ion source parts in CI ion source
•
Air leaks in reagent gas flow path
•
CI filament has stretched and sagged:
•
High EMV
•
Linear (no inflection point) electron energy (EIEnrgy) ramp
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......