CI Troubleshooting
3
77
No or low reagent gas signal
If you have just installed the CI ion source and have an air leak or large amounts
of water in the system after running one or more autotunes, the ion source is
probably dirty.
Fix the air leak. Clean the ion source, and bake out for 2 hours before tuning.
See the
Operation Manual
.
The wrong reagent gas is flowing.
Turn on the correct reagent gas for your tune file.
Ion polarity is set to Negative. No reagent gas ions are formed in NCI.
Switch to Positive ionization mode.
The reagent gas flow is set too low.
Increase the reagent gas flow.
Reagent gas supply tubing is blocked, kinked, pinched, or disconnected.
Inspect and, if necessary, repair or replace the reagent gas supply tubing.
Wrong filament wires are connected to filament.
Ensure that the filament 1 wires are connected to the CI ion source filament,
and that the filament 2 wires are connected to the dummy filament.
Carbon has built up on the filament, or the filament has sagged out of alignment.
Inspect the filament, if necessary, replace the filament.
Too much air or water in the system.
Run the methane pretune. Peaks at
m/z
32 and 19 usually indicate air and
water, respectively. Bake out and purge the instrument until there is no visible
peak at
m/z
32, and the peak at
m/z
19 is reduced to a very low level. If the
peak at
m/z
32 does not decrease, an air leak is likely. See
page 66 for more information.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......