5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
6
Place a column nut with a blank ferrule on the end of the interface. This
will help keep contamination out of the MSD.
7
Disconnect the GC/MSD interface cable and the LVDS cable (if applicable).
Disconnecting either cable with the GC on can cause a fault condition.
8
Continue to move the MSD until you have access to the part requiring
maintenance.