118
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Remove the DP
Materials needed
•
Aluminum foil, clean
•
Gloves, oil-resistant
Procedure
1
Vent the MSD.
2
Separate the MSD from the GC. (See
“To Separate the MSD from the 7890
3
Disconnect high vacuum power (HIVAC POWER) cable from the back panel
of the MSD. (This is the thick black cable that emerges near the bottom of
the pump.)
4
Disconnect the DP temperature sensor wires from the wiring harness.
5
Support the DP with one hand.
WA R N I N G
Treat the DP fluid as hazardous, as it may contain traces of toxic chemicals.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......