General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
105
To Move or Store the MSD when Connected to a 7890 GC
Materials needed
•
Ferrule, blank (5181-3308)
•
Interface column nut (05988-20066)
•
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
Procedure
1
Move the MSD away from the GC. (See
2
Tighten the vent valve.
3
Install the interface nut with the blank ferrule.
4
Open the analyzer cover.
5
Finger-tighten the side plate thumbscrews. (See
6
Plug the MSD power cord in.
7
Switch the MSD on for 5 minutes to establish a rough vacuum.
8
Switch the MSD off.
9
Close the analyzer cover.
10
Disconnect the LAN, remote, and power cables.
C A U T I O N
Do not overtighten the side plate thumbscrews. Overtightening will strip the threads in
the analyzer chamber. It will also warp the side plate and cause leaks.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......