114
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Install the Exhaust Filter on the IDP3 Dry Pump
Materials needed
•
Exhaust filter (G3170-80041)
Procedure
1
Disconnect the foreline pump exhaust hose from the adapter.
2
Unscrew the adapter from the pump.
3
Screw the filter on to the pump.
Figure 12
IDP3 pump with exhaust filter
Exhaust filter
Exhaust hose
Hose adapter