![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 66](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608066.webp)
66
3
CI Troubleshooting
Air Leaks
How do I know if I have an air leak?
Large
air leaks can be detected by vacuum symptoms: loud gurgling noise
from the foreline pump, inability of the turbo pump to reach 95% speed, or, in
the case of smaller leaks, high pressure readings on the high vacuum gauge
controller.
The mass flow controller is calibrated for methane, and the high vacuum gauge
controller is calibrated for nitrogen, so measurements are not accurate in
absolute terms.
Familiarize yourself with the measurements on
your
system under operating
conditions. Watch for
changes
that may indicate a vacuum or gas flow
problem.
Always
look for small air leaks when setting up methane flow. Run the
methane
pretune, starting with a good PCI tune file (
on page 67). The
abundance of
m/z
19 (protonated water) should be less than 50% of
m/z
17 for
acceptable PCI performance. For NCI, the abundance of
m/z
19 (protonated
water) should be less than 25% that of
m/z
17. If the MSD was just pumped
down, look for the abundance of
m/z
19 to be decreasing.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......