200
5977B Series MSD Troubleshooting and Maintenance Manual
7
Analyzer
Filament care
Similar to the filaments in incandescent light bulbs, the ion source filaments
will eventually burn out. Certain practices reduce the chance of early failure:
•
If you have an optional G3397B Micro-Ion vacuum gauge, use it to verify
that the system has an adequate vacuum before turning on the analyzer,
especially after any maintenance was performed.
•
If you are controlling your MSD from the
Manual Tune
screen, always select
MSOff
before changing any of the filament parameters.
•
When setting up data acquisition parameters, set the solvent delay so that
the analyzer will
not
turn on while the solvent peak is eluting.
•
When the software prompts
Override solvent delay
?
at the beginning of a run,
always
select
NO
.
•
Higher emission current reduces filament life.
•
Higher electron energy reduces filament life.
•
Leaving the filament on for short times (
1 minute) during data acquisition
reduces filament life.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......