5977B Series MSD Troubleshooting and Maintenance Manual
6
Vacuum System
The diffusion pump transports gas by momentum transfer. The heater boils a
special fluid (a polyphenyl ether) inside the stack. As the vapor pressure
increases, the pump fluid vapor is forced out and downward through nozzles
in the stack. The vapor forced out of these nozzles strikes the gas molecules
that are present. This forces the gas molecules down toward the outlet near
the bottom of the pump. Another nozzle in the stack points directly at the
outlet, and forces the gas molecules out. The vapor condenses on the sides of
the pump, and the liquid drains down to the bottom. The liquid is boiled again,
and is reused continuously.
Figure 34
Diffusion pump parts
Foreline gauge assembly
Foreline gauge cable – part of
high vacuum control cable
Diffusion pump outlet
KF10/16 seal
Foreline hose
and hose clamp
KF10/16 clamp
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......