130
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Purge the Calibration Valves
EI calibration valve
After adding new PFTBA to the calibrant vial, you must purge the air out of
the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
In
Tune and Vacuum Control
view, select
Purge Calibrant Valve
under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
CI calibration valve
After adding new PFDTD to the calibrant vial, you must purge the air out of
the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
Verify that
PCICH4.U
is loaded.
3
In
Tune and Vacuum Control
view, select
Purge Calibrant Valve
under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
C A U T I O N
After removing a calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in damage to the filaments and the electron multiplier.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......