CI Troubleshooting
3
71
Poor vacuum without reagent gas flow
Excess water in the background
Scan from 10 to 40
m/z
. A large peak at
m/z
19 (>
m/z
17) indicates water in the
background. If water is present, allow the instrument to bake out more, and
flow reagent gas through the lines to purge any accumulated water.
Air leak
Run Methane Pretune. See the
Operation Manual
. A visible peak at
m/z
32
indicates air in the system. Check for, and correct any leaks. See the
Leaks
section at the beginning of this chapter.
The foreline pump is not working properly
For the standard foreline pump, replace the pump oil. If that does not help, or
for the dry foreline pump, it may be necessary to replace the pump. Contact
your local Agilent Technologies Customer Engineer.
The turbo pump is not working properly
Check the pump speed. It should be at least 95%. Contact your local Agilent
Technologies service representative.
C A U T I O N
Use of ammonia as reagent gas can shorten the life of the foreline pump oil (with
standard pump) and possibly of the foreline pump itself.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......