Place a digital multimeter between B11-Tp2 (+)
and B11-Tp6 (GND).
Control Snapshot and check that the voltage
read on the multimeter is 800mV (-0 +50mV).
After performing the two adjustments described
above, control Snapshot and note down the
voltage present between OFC01-Tp3 (+) and
OFC01-Tp4 (GND) (
“point B
”, about –3,87V).
In Manual Fluoroscopy mode (without x-rays),
adjust the potentiometer OFC01-P2 to read
between OFC01-Tp3 (+) and OFC01-Tp4
(GND), a voltage equal to the value read in
“point A”
.
In Manual Fluoroscopy mode (without x-rays),
enter Setup mode (code 1001), on the board
OFC01 the led Ld1 (yellow) lights up, adjust the
potentiometer OFC01-P1 to read between
OFC01-Tp3 (+) and OFC01-Tp4 (GND), a
voltage equal to the value read in
“point B”
.
Repeat more times the adjustment because the
two potentiometers OFC01-P1 and OFC01- P2
influence each other.
Posizionare un multimetro digitale tra B11-Tp2
(+) e B11-Tp6 (GND).
Comandare Snapshot e verificare che la
tensione letta sul multimetro sia 800mV (-0
+50mV).
Eseguite le due regolazioni sopra descritte,
comandare Snapshot e prendere nota della
tensione presente tra OFC01-Tp3 (+) e OFC01-
Tp4 (GND) (
“punto B
”, circa –3,87V).
In modalità Scopia Manuale (senza Rx),
regolare il potenziometro OFC01-P2 per
leggere tra OFC01-Tp3 (+) e OFC01-Tp4
(GND), una tensione pari al valore letto in
“punto A”
.
In modalità Scopia Manuale (senza Rx), entrare
in modalità Setup (codice 1001), sulla scheda
OFC01 si accende il led Ld1 (giallo), regolare il
potenziometro OFC01-P1 per leggere tra
OFC01-Tp3 (+) e OFC01-Tp4 (GND), una
tensione pari al valore letto in
“punto B”
.
Ripetere più volte la regolazione perché i due
potenziometri OFC01-P1 e OFC01-P2 si
influenzano a vicenda.
Fig. 24
Board OFC01
Содержание ARCOVIS 3000 R
Страница 153: ...Villa Sistemi Medicali ARCOVIS 30000 S R Adjustments Regolazioni 4 File 201180 B 00 20 doc ADJUSTMENTS REGOLAZIONI 4...
Страница 196: ...Villa Sistemi Medicali ARCOVIS 3000 S R Faults Finding 5 File 201181 A 00 20 doc FAULTS FINDING RICERCA GUASTI 5...
Страница 350: ......
Страница 352: ...Villa Sistemi Medicali ARCOVIS 3000 S R Parts List 7 File 201183 E 00 20 doc PARTS LIST 7...
Страница 353: ...Parts List 7 ARCOVIS 3000 S R Villa Sistemi Medicali File 201183 E 00 20 doc This page is intentionally left blank...
Страница 381: ......
Страница 382: ......
Страница 383: ......
Страница 384: ......
Страница 385: ......
Страница 386: ......
Страница 387: ......
Страница 388: ......
Страница 389: ......
Страница 390: ......
Страница 391: ......
Страница 392: ......
Страница 393: ......
Страница 394: ......
Страница 395: ......
Страница 396: ......
Страница 397: ......
Страница 398: ......
Страница 399: ......
Страница 400: ......
Страница 401: ......
Страница 402: ......
Страница 403: ......
Страница 404: ......
Страница 405: ......
Страница 406: ......
Страница 407: ......
Страница 408: ......
Страница 409: ......
Страница 410: ......
Страница 411: ......
Страница 412: ......
Страница 413: ......
Страница 414: ......
Страница 415: ......
Страница 416: ......
Страница 417: ......
Страница 418: ......
Страница 419: ......
Страница 420: ......
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......