Villa Sistemi Medicali
ARCOVIS 30000 S/R – Adjustments / Regolazioni - 4
[File: 201180-B-00-20.doc]
Rev.B - Page 7/40
2.2.2. Tube calibration in radiography / Calibrazione del tubo in grafia
Place the oscilloscope on the board PSM18:
channel A: probe Tp13 (format 1V = 10mA) –
Tp3 GND.
channel B: probe Tp14 (format 1V = 20kV) –
Tp3 GND.
Place the digital multimeter on:
B8-Tp6 (probe, format 1V-100mA) and B8-Tp1
(GND).
Posizionare l’oscilloscopio sulla scheda
PSM18:
canale A: sonda Tp13 (formato 1V = 10mA) –
Tp3 GND.
canale B: sonda Tp14 (formato 1V = 20kV) –
Tp3 GND.
Posizionare il multimetro digitale su:
B8-Tp6 (sonda, formato 1V-100mA) e B8-Tp1
(GND).
Perform a series of three exposures in
radiography (60kV, 80kV, 100kV 10mAs).
Check that the mA wave-form is as straight as
possible (Fig.1 – B).
In case the wave-form is not straight (Fig.1 –
A), enter Setup mode and modify the value
“ADJ START IFIL RAD” to get a wave-form
similar to the one reported in Fig.1 - B.
Eseguire una serie di tre esposizioni in grafia
(60kV, 80kV, 100kV 10mAs).
Verificare che la forma d’onda dei mA sia il più
rettilinea possibile (Fig.1 – B).
Se la forma d’onda non è rettilinea (Fig.1 – A),
entrare in modalità Setup e modificare il valore
“ADJ START IFIL RAD” per ottenere una forma
d’onda simile a quella riportata in Fig.1 - B.
Fig. 1 - A
Fig. 1 - B
Set the kV values in the “kV” column of the
table 5. Select radiography, measure the
values of kV and mA and check that they are
included into the reported tolerances.
Impostare i valori di kV riportati nella colonna
“kV” della tab. 5. Selezionare Grafia, rilevare i
valori di kV e mA e verificare che rientrino nelle
tolleranze riportate.
Tab. 5 – Radiography
kV
kV ±5%
mAs
mA
mA ±7,5%
40
38.0 ÷ 42.0
100
23.1 ÷ 26.8
50
47.5 ÷ 52.5
100
23.1 ÷ 26.8
60
57.0 ÷ 63.0
80
23.1 ÷ 26.8
70
66.5 ÷ 73.5
80
23.1 ÷ 26.8
80
76.0 ÷ 84.0
80
23.1 ÷ 26.8
90
85.5 ÷ 94.5
63
23.1 ÷ 26.8
100 95.5
÷105.5 63
25
23.1 ÷ 26.8
110
104.5 ÷115.5
63
22,7
21.3 ÷ 24.7
120
114 ÷ 126
63
20,8
18.5
÷ 21.5
Содержание ARCOVIS 3000 R
Страница 153: ...Villa Sistemi Medicali ARCOVIS 30000 S R Adjustments Regolazioni 4 File 201180 B 00 20 doc ADJUSTMENTS REGOLAZIONI 4...
Страница 196: ...Villa Sistemi Medicali ARCOVIS 3000 S R Faults Finding 5 File 201181 A 00 20 doc FAULTS FINDING RICERCA GUASTI 5...
Страница 350: ......
Страница 352: ...Villa Sistemi Medicali ARCOVIS 3000 S R Parts List 7 File 201183 E 00 20 doc PARTS LIST 7...
Страница 353: ...Parts List 7 ARCOVIS 3000 S R Villa Sistemi Medicali File 201183 E 00 20 doc This page is intentionally left blank...
Страница 381: ......
Страница 382: ......
Страница 383: ......
Страница 384: ......
Страница 385: ......
Страница 386: ......
Страница 387: ......
Страница 388: ......
Страница 389: ......
Страница 390: ......
Страница 391: ......
Страница 392: ......
Страница 393: ......
Страница 394: ......
Страница 395: ......
Страница 396: ......
Страница 397: ......
Страница 398: ......
Страница 399: ......
Страница 400: ......
Страница 401: ......
Страница 402: ......
Страница 403: ......
Страница 404: ......
Страница 405: ......
Страница 406: ......
Страница 407: ......
Страница 408: ......
Страница 409: ......
Страница 410: ......
Страница 411: ......
Страница 412: ......
Страница 413: ......
Страница 414: ......
Страница 415: ......
Страница 416: ......
Страница 417: ......
Страница 418: ......
Страница 419: ......
Страница 420: ......
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......