Choose the available APR programs and
repeat the measurement of the mA values for
every kV decade.
Scegliere i programmi APR disponibili e ripetere
le rilevazioni dei valori di mA per ogni decade di
kV.
Tab. 3 – Continuous APR fluoroscopy
MA_LOW
mA_LUNG
mA-HIGH (*)
kV
kV ±5%
mA
mA ±7,5%
mA
mA ±7,5%
mA
mA ±7,5%
40
38.0÷42.0 0.70 0.65
÷
0.75
0.50
0.46 ÷ 0.54
1
0.92
÷
1.08
50
47.5÷52.5 2.40 2.22
÷
2.58
1.00
0.92 ÷ 1.08
4
3.70
÷
4.30
60
57.0÷63.0 4.20 3.88
÷
4.52
3.00
2.77 ÷ 3.23
10
9.25
÷
10.75
70
66.5÷73.5 5.60 5.18
÷
6.02
5.00
4.62 ÷ 5.38
15
13.87
÷
16.13
80
76.0÷84.0 6.40 5.92
÷
6.88
7.00
6.47 ÷ 7.53
13
12.02
÷
13.98
90
85.5÷94.5 6.40 5.92
÷
6.88
6.50
6.01 ÷ 6.99
12
11.10
÷
12.90
100
95.5÷105.5 6.40 5.92
÷
6.88
6.25
5.78 ÷ 6.72
11
10.17
÷
11.83
110
104.5÷115.5 6.60
6.10
÷
7.10
6.00
5.55 ÷ 6.45
10
9.25
÷
10.75
120
114 ÷ 126
6.60
6.10
÷
7.10
5.50
5.09
÷
5.91
9
8.32
÷
9.68
(*) The mA-HIGH curve is available only with
rotating anode, 1K camera and memory of
HRC1000 type.
(*) La curva mA-HIGH è disponibile solo con
anodo rotante, camera 1K e memoria tipo
HRC1000.
Select the following programs and repeat the
mA values monitoring for each marked kV
decade.
Not all the programs are available for the
different configurations.
Scegliere i programmi seguenti e ripetere le
rilevazioni dei valori di mA per ogni decade di
kV segnate.
Non tutti i programmi sono disponibili per le
diverse configurazioni.
Tab. 4 – Digital Radiography “SnapShot“
kV
kV ±5%
Fixed Anode
0,5K camera
Rotating Anode
0,5k camera
Rotating Anode
1k camera
Rotating Anode
1K camera HRP
memory
mA
mA ±7,5%
mA
mA ±7,5%
mA
mA ±7,5%
mA
mA ±7,5%
40
38.0÷42.0 1.00 0.92
÷
1.07
1.00
0.92 ÷ 1.07
2,5
2.31
÷
2.69
60
55.20
÷
64.50
70
66.5÷73.5 10.0 9.25
÷
10.75
12.00
11.1 ÷ 12.9
---
---
---
---
80
76.0÷84.0 10.0
9.25
÷
10.75
12.00
11.1 ÷ 12.9
30
27.75
÷
32.25
60
55.20
÷
64.50
110
104.5÷115.5 8.00 7.40
÷
8.60
--- --- --- --- --- ---
120
114 ÷ 126
---
---
9.00
8.32
÷
9.68
10
9.25
÷
10.75
60
55.20
÷
64.50
Tab. 5 – HRP memory
HCF Fluoroscopy
Pulsed Radiography
Pulsed Radiography
kV
kV ±5%
mA
mA ±7,5%
mA
mA ±7,5%
mA
mA ±7,5%
40
38.0÷42.0 30.0 27.75
÷
32.25
60.0
55.5 ÷ 64.5
45
41.62
÷
48.38
80
76.0÷84.0 30.0 27.75
÷
32.25
60.0
55.5 ÷ 64.5
45
41.62
÷
48.38
120
114 ÷ 126
30.0
27.75
÷
32.25
60.0
55.5 ÷ 64.5
45
41.62
÷
48.38
Содержание ARCOVIS 3000 R
Страница 153: ...Villa Sistemi Medicali ARCOVIS 30000 S R Adjustments Regolazioni 4 File 201180 B 00 20 doc ADJUSTMENTS REGOLAZIONI 4...
Страница 196: ...Villa Sistemi Medicali ARCOVIS 3000 S R Faults Finding 5 File 201181 A 00 20 doc FAULTS FINDING RICERCA GUASTI 5...
Страница 350: ......
Страница 352: ...Villa Sistemi Medicali ARCOVIS 3000 S R Parts List 7 File 201183 E 00 20 doc PARTS LIST 7...
Страница 353: ...Parts List 7 ARCOVIS 3000 S R Villa Sistemi Medicali File 201183 E 00 20 doc This page is intentionally left blank...
Страница 381: ......
Страница 382: ......
Страница 383: ......
Страница 384: ......
Страница 385: ......
Страница 386: ......
Страница 387: ......
Страница 388: ......
Страница 389: ......
Страница 390: ......
Страница 391: ......
Страница 392: ......
Страница 393: ......
Страница 394: ......
Страница 395: ......
Страница 396: ......
Страница 397: ......
Страница 398: ......
Страница 399: ......
Страница 400: ......
Страница 401: ......
Страница 402: ......
Страница 403: ......
Страница 404: ......
Страница 405: ......
Страница 406: ......
Страница 407: ......
Страница 408: ......
Страница 409: ......
Страница 410: ......
Страница 411: ......
Страница 412: ......
Страница 413: ......
Страница 414: ......
Страница 415: ......
Страница 416: ......
Страница 417: ......
Страница 418: ......
Страница 419: ......
Страница 420: ......
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......