Villa Sistemi Medicali
ARCOVIS 30000 S/R – Adjustments / Regolazioni - 4
[File: 201180-B-00-20.doc]
Rev.B - Page 13/40
0,5K
2
SYSTEM VIDEO ADJUSTMENTS / REGOLAZIONI VIDEO SISTEMA 0,5K
2
3.3.3. Collimator fields adjustment / Regolazione
campi collimatore
Enter Service Mode, set the password “1001” to
calibrate the collimator positions. Adjust the
collimator openings (iris and shutters) for each
one of the three intensifier fields.
If it is not specified differently, the shutters and
the iris must be outside the visible field.
At the end of the adjustment, exit from Service
Mode and, on request, store the data.
Entrare in modalità Service Mode, impostare la
password “1001” per calibrare le posizioni del
collimatore. Regolare le aperture del collimatore
(iride e lamelle) per ognuno dei tre campi
dell’intensificatore.
Se non diversamente specificato, le lamelle e
l’iride devono trovarsi al di fuori del campo
visibile.
Alla fine della regolazione uscire da Service
Mode e, alla richiesta, salvare i dati.
3.4. Dimension of automatic kV control area / Dimensione area di controllo
kV automatici.
Ref. Fig. 7
Set the min. kV and mA data in manual
fluoroscopy.
Move the Jumper N23_ANB1-J2 in the position
2-3 "TEST AUTO CIRCLE".
Control x-rays and at the same time adjust the
potentiometer N23_ANB1-P5 (AUTO CIRCLE)
to get on the monitor a circle of:
10cm ±0,5
for normal applications
8cm ±0,5
for applications for lithotripsy
Place again N23_ANB1-J2 in the position 1-2.
Impostare i dati minimi di kV e mA in
fluoroscopia manuale.
Spostare il Jumper N23_ANB1-J2 nella
posizione 2-3 "TEST AUTO CIRCLE".
Comandare RX e contemporaneamente
regolare il potenziometro N23_ANB1-P5 (AUTO
CIRCLE), per ottenere sul monitor un cerchio di:
10cm ±0,5
per applicazioni normali
8cm
±0,5 per applicazioni per
litotripsia
Riposizionare N23_ANB1-J2 nella posizione 1-
2.
R94
R157
J5
N23_ANB1
P4
P11
P7
P10
P9
P8
P6
P5
J4
1 2 3
P1
P2
P3
J2
1 2 3 4
SW4
1 2 3
J6
1 2 3
SW2
1
2
3
R156
R163
SW3
1
2
3
1
2
3
Fig. 7
Содержание ARCOVIS 3000 R
Страница 153: ...Villa Sistemi Medicali ARCOVIS 30000 S R Adjustments Regolazioni 4 File 201180 B 00 20 doc ADJUSTMENTS REGOLAZIONI 4...
Страница 196: ...Villa Sistemi Medicali ARCOVIS 3000 S R Faults Finding 5 File 201181 A 00 20 doc FAULTS FINDING RICERCA GUASTI 5...
Страница 350: ......
Страница 352: ...Villa Sistemi Medicali ARCOVIS 3000 S R Parts List 7 File 201183 E 00 20 doc PARTS LIST 7...
Страница 353: ...Parts List 7 ARCOVIS 3000 S R Villa Sistemi Medicali File 201183 E 00 20 doc This page is intentionally left blank...
Страница 381: ......
Страница 382: ......
Страница 383: ......
Страница 384: ......
Страница 385: ......
Страница 386: ......
Страница 387: ......
Страница 388: ......
Страница 389: ......
Страница 390: ......
Страница 391: ......
Страница 392: ......
Страница 393: ......
Страница 394: ......
Страница 395: ......
Страница 396: ......
Страница 397: ......
Страница 398: ......
Страница 399: ......
Страница 400: ......
Страница 401: ......
Страница 402: ......
Страница 403: ......
Страница 404: ......
Страница 405: ......
Страница 406: ......
Страница 407: ......
Страница 408: ......
Страница 409: ......
Страница 410: ......
Страница 411: ......
Страница 412: ......
Страница 413: ......
Страница 414: ......
Страница 415: ......
Страница 416: ......
Страница 417: ......
Страница 418: ......
Страница 419: ......
Страница 420: ......
Страница 421: ......
Страница 422: ......
Страница 423: ......
Страница 424: ......