Appendix A
Specifications
Jitter and Wander specifications
T-BERD/MTS/SC Getting Started Guide
March 2018
22046537, Rev. 016
Page 235
•
Structured test signals (for SDH as described in ITU-T Recommendation O.172)
•
Sine wave modulation
•
Standard filters: Highband
Measurement error
The stated measurement error applies without restriction to standard range, for values
>25 UIp in extended range.
•
Maximum measurement error: ± R % of reading ± W
The value for W (fixed error) is taken from
. The value for R (variable error)
can be taken from
ITU-T Recommendations O.172 (04/2005) and O.173 (03/2003)
as
applicable.
RMS jitter
This section describes the RMS jitter specifications.
Range and resolution
describes the RMS range and resolution.
Measurement accuracy
describes the RMS jitter measurement accuracy.
Table 159
W (fixed error)
Range
Wide-band Filter
Extended-band Filter
W in UIpp
W in UIpp
Standard range
0.035
1
1. Mapping SDH VC-4/SONET STS-1, payload pattern PRBS31, environmental tempe20
°
C to +30
°
C.
0.5
Table 160
RMS range and resolution
(peak-peak)
Standard range
Extended range
RMS range
0 to 25 UI
0 to 400 UI
Resolution
0.0001 UI
0.01 UI
Table 161
RMS jitter measurement accuracy
Item
Description
Standard range
± R% of measured value ± 0.003 UI
1
1. The value R (variable error) is taken from
ITU-T Recommendation O.172 (04/2005) and O.173 (03/2003)
as appli-
cable.
Extended range
± R% of measured value ± 0.05 UI
Integration time
1 second