Chapter 7
Optical Tools
Inspecting fiber
T-BERD/MTS/SC Getting Started Guide
March 2018
22046537, Rev. 016
Page 159
2
Select the
MicroScope
icon.
The Fiber Microscope screen appears.
3
Connect the microscope to the optical patch cord or bulkhead.
The video of the fiber appears.
4
Specify which
Tip
is being used on the microscope.
5
Adjust the focus as needed using the wheel on the microscope. The color of the
vertical bar on the left indicates the quality of the focus: green is good, yellow is
ok, red is poor. Older microscopes have a blue bar only.
Optional.
To verify which version you have, use the Viavi
FiberChek Pro
desktop
application.
6
Select the inspection
Profile
.
If you have a custom profile, do one of the following:
–
Use the file manager to put the file into the disk/bert/scope/profiles/ folder,
and it will appear in the P5000i viewer the next time it is launched (if it is
already running, it will need to be closed and reopened).
–
Click the
Import from USB
button next to the profile list to import your
custom profile. Once imported it may be selected in the viewer.
7
Select
Other settings
and then check the
Auto-center
check box. If checked, the
live picture automatically centers on the fiber center. If not checked, the picture
will center at the last location of a fiber center following an analysis. High magnifi-
cation is always automatically centered.
8
To manually inspect the fiber, do the following:
a
Observe the video of the fiber, and if desired, magnify the image by selecting
the
Zoom In
button.
b
Select
Other settings
and then specify
Freezes
as the
Test Button
setting.
This captures a snapshot.
c
Select the
Test
button.
d
To compare the current capture to the live feed, select
Overlay
e
To return to the live video, select the
Live
button.
9
To test the fiber, do the following:
a
Select
Other settings
and then specify
Tests
as the
Test Button
setting.
b
Select the
Test
button.
NOTE:
Although some microscopes can inspect multi-fiber or ribbon fiber (depending
on the microscope and the tip used) the T-BERD
⁄
MTS 5800’s microscope
application supports simplex fiber only.
NOTE:
The test will fail if the focus quality is poor (vertical bar is red).