Chapter 16
Performing RFoCPRI Testing
JD700B Series User’s Guide
551
RAI:
Remote Alarm Indication (any errors including LOS and LOF)
SDI:
SAP Defect Indication (link not to be used for serivce access points)
Optic Rx Level:
If the optical signal level is lower than the threshold, the CPRI link may not
work properly and the RRH may not be in service mode.
2
Optional
. You can use the
Save Logging Data
soft key to record the event logging in
.csv
file
format. See “Using save” on page 35 for more information.
3
To clear the previous event history and start a new monitoring, press the
MEASURE SETUP
hot
key and then the
Clear History
soft key.
Measurement example
Figure 336 CPRI layer 2 monitoring
Performing layer 2 term testing
The out-of-service Layer 2 Term is to test CPRI fronthaul in the terminated state and it is used to monitor link
maintenance alarms of the physical layer (Layer 1) that is delivered on the L1 in-band protocol, which
resides on the layer 2 of CPRI. It also monitors the received optical power level.
In addition, CellAdvisor JD700B series can emulate the Base Band Unit (BBU) without C&M messages.
Once the RRH enters the operation state, JD700B series can generate an alarm or error and send it to the
RRH so that you can verify the optical cabling and proper RRH operation at the ground.
Connecting cables
In the terminate mode, you can perform the out-of-service testing. Before starting the test, you need to
connect cables with a tap or without a tap as illustrated in
A SFP/SFP+ transceiver that is connected to your JD700B series must be compatible with your DUT and
you must have your module information such as line rate, wavelength, and mode (MM or SM) handy.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...