Chapter 4
Using Spectrum Analyzer
70
JD700B Series User’s Guide
Optional
.
To define the amount of channel increment for the rotary knob:
1
Press the
Channel Step
soft key, if the
Channel
is selected.
2
Enter a value by using the numeric keys.
3
Press the
Enter
soft key.
NOTE
This setting option is not available in the Spurious Emissions mode.
Setting frequency for 28 GHz downconverter
Once the CellAdvisor is connect to 28 GHz downconverter, the center frequency is automatically set to 28
GHz. You can measure the center frequency ranged from 27 to 30 GHz with the span or the start/stop
frequencies. You cannot set the frequency with the channel number and span when the 28 GHz
downconverter is in operation.
Procedure
To set the center frequency and span:
1
Press the
FREQ/DIST
hard key.
2
Press the
Center Frequency
soft key.
3
Enter a value from 27 to 30 GHz by using the numeric keys, the rotary knob, or the arrow keys.
4
Select the unit as
GHz
among the 4 options:
GHz
,
MHz
,
kHz
, or
Hz
.
5
Press the
Span
soft key, and then do one of the following:
To set the span manually, select
Span
, enter a value, and then select the unit.
To set the span automatically, select
Full Span
,
Zero Span
, or
Last Span
as needed.
NOTE
The
Zero Span
option is available only in the Spectrum mode. If you have selected
Zero
Span
, select
TRIGGER > Free Run
,
External
,
GPS
,
Video
, or
Display Position
.
Optional
.
To define the amount of frequency increment for the rotary knob:
1
Press the
Frequency Step
soft key, if the
Freq
is selected.
2
Enter a value by using the numeric keys.
3
Press the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
To set the start and stop frequencies:
1
Press the
FREQ/DIST
hard key.
2
Press the
Start Frequency
soft key.
3
Enter a value from 27 to 30 GHz by using the numeric keys, the rotary knob, or the arrow keys.
4
Select the unit as
GHz
among the 4 options:
GHz
,
MHz
,
kHz
, or
Hz
.
5
Press the
Stop Frequency
soft key.
6
Enter a value from 27 to 30 GHz by using the numeric keys, the rotary knob, or the arrow keys.
7
Select the unit as
GHz
among the 4 options:
GHz
,
MHz
,
kHz
, or
Hz
.
Setting amplitude
Reference level and attenuation
You can set the reference and attenuation levels automatically or manually to optimize the display of the
traces measured, as you desire.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...