Chapter 8
Using LTE-FDD Signal Analyzer
JD700B Series User’s Guide
225
NOTE
Selecting
Channel
sets the standard channel to
Band Global
automatically so that you
can set the
Channel Number
without choosing a standard channel band from the list.
Press the
Lowest Channel
soft key to set the starting channel.
Enter a value by using the numeric keys. You can also use the rotary knob.
Press the
Enter
soft key.
Press the
Highest Channel
soft key to set the stopping channel.
Enter a value by using the numeric keys. You can also use the rotary knob.
Press the
Enter
soft key.
Measurement example
Figure 120 Multi-ACLR measurement with LTE-FDD signal analyzer
NOTE
You can use the
LIMIT
hot key to analyze your measurements with the user-definable
limit and Pass/Fail indication. See “Setting limit for RF tests” on page 227 for more
information.
Spurious emissions
Out-of-band emissions are unwanted emissions immediately outside the channel bandwidth resulting from
the modulation process and non-linearity in the transmitter but excluding spurious emissions. The Spurious
Emissions measurement is to identify and determine the power level of out-of-band spurious emission within
the necessary channel bandwidth and modulated signal measured at the RF port of the Base Station.
Setting measure setup
After configuring test parameters as described in the “Configuring test parameters” on page 214, you can
continue your measurement.
Procedure
1
Press the
MEASURE SETUP
hot key.
2
To set up the range table and parameters, press the
Range Table
soft key and then complete
the following steps:
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...