JD700B Series User’s Guide
Figure 464 EMF spectrum measurement with EMF analyzer
Setting trace
You can display up to six traces on the measurement chart simultaneously.
Procedure
1
Press the
TRACE/DISPLAY
hard key.
2
Press the
Select Trace
soft key, and then select the trace number from the following choices:
T1
: Current EMF Trace
T2
: EMF Trace of X-axis
T3
: Isotropic EMF Trace
T4
: EMF Trace of Y-axis
T5
: Accumulated Isotropic EMF Trace
T6
: EMF Trace of Z-axis
The legend shape of the selected trace changes from square to round to indicate that the trace
is the active one now.
3
Toggle the
Trace View
soft key to
On
to show the selected trace on the display or to
Off
to hide it
from the display.
Setting limit
You can set three different limits to analyze the results, as you desire.
Procedure
To set the display line:
1
Press the
LIMIT
hot key.
2
Toggle the
Display Line
soft key to
On
.
The straight horizontal line appears to be used as a visual reference only.
3
Press the
Display Line
soft key, and then enter a value that you want to set as a reference.
4
Select your desired unit soft key from the available choices, which vary depending on your
selection of the unit in the
Units
setting in Amp/Scale.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...