JD700B Series User’s Guide
765
Figure 463 Connection using an isotropic antenna
CAUTION
The maximum power for the
Spectrum Analyzer RF In
port is +25 dBm (0.316 W) for
JD780B series and +20 dBm (0.1 W) for JD740B series.
Selecting measurement mode
Procedure
1
Press the
MODE
hard key.
2
Press the
More (1/2)
soft key and then the
EMF Analyzer
soft key.
This menu is activated only after you have not installed the license on your CellAdvisor.
3
Press the
MEASURE
hot key, and then select the measurement mode from the following
choices:
EMF Spectrum
EMF Replayer
Configuring test parameters
Setting frequency
You need to set the frequency range to be measured with either the center frequency/span or the start/stop
frequencies. You can also set the frequency with the channel number and span.
Procedure
To set the center frequency and span:
1
Press the
FREQ/DIST
hard key.
2
Toggle the
Unit
soft key and select
Freq
.
3
To set the center frequency, complete the following steps:
Press the
Center Frequency
soft key.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...