Chapter 10
Using CDMA Signal Analyzer
356
JD700B Series User’s Guide
Procedure
1
Press the
TRIGGER
hot key.
2
Select the trigger source option from the choices:
Internal
,
External
, and
GPS
.
Setting external clock
To enhance the reliability of modulation analysis measurements the JD700B series must be synchronized
with a base station. When an external clock is not supplied, the instrument works with its built-in internal
high-accuracy time base and some measurement results may exhibit inaccurate values. Therefore, it is
highly recommended that you use the same reference clock as the signal source. You can use the
TRIGGER
hot key to set the external clock.
Table 15
Standards of external clock operation
Clock
Type
Reference
Port
Switching Standard
External Reference Internal
Internal 10 MHz
Default
External
External Ref Clock
Ext Ref (SMA)
Manual switching
(Input signal level >-3 dBm)
GPS
GPS Clock
GPS (SMA)
Automatic switching
(GPS signal received)
Trigger
Internal
Internal Trigger
Default
External
External Trigger
Trigger (SMA)
Manual switching
GPS
GPS
GPS (SMA)
Manual switching
Figure 208 Connection ports for external reference clock
Procedure
1
Connect an external reference or a GPS antenna to the JD700B series.
2
Press the
SYSTEM
hot key.
3
Press the
Freq Reference
soft key, and then select the reference option:
Internal
,
External 10
MHz
,
External 13 MHz
,
External 15 MHz
, or
GPS
.
NOTE
When a GPS antenna is connected and locked, GPS coordinates (longitude and latitude)
are displayed on the screen and frequency reference is automatically set to GPS in the
System > Freq Reference
.
Содержание CellAdvisor JD700B Series
Страница 2: ...CellAdvisorTM JD700B Series FW3 120 User s Guide ...
Страница 105: ...Chapter 4 Using Spectrum Analyzer 104 JD700B Series User s Guide Figure 44 PIM detection for multiple carriers ...
Страница 406: ...Chapter 11 Using EV DO Signal Analyzer JD700B Series User s Guide 405 3 Pilot 4 MAC 64 5 MAC 128 ...
Страница 734: ...Chapter 18 Perfroming BBU Emulation JD700B Series User s Guide 733 Figure 441 Example of verification of link status ...
Страница 763: ...Chapter 21 Performing Fiber Inspection 762 JD700B Series User s Guide Figure 461 Fiber inspection report ...
Страница 808: ...Appendix JD700B Series User s Guide 807 JD740B series ...
Страница 811: ...Appendix 810 JD700B Series User s Guide Figure 470 Append keywords window ...